OPERATION PROCEDURES FOR PHILIPS FEI CM 20 TRANSMISSION ELECTRON MICROSCOPE General Comments Normally the microscope will be left in the power ON standby ready to operate condition i e all circuits will be continuously energized and the microscope will be under full vacuum MAIN VACUUM HIGH TENSION lamps on the right control panel are all lit if the PANEL DIM is pressed If it is not in this condition do not operate the microscope Seek help from authorized personnel If the microscope is in standby ready condition proceed with the following abbreviated procedures for high voltage application and alignment before beginning any observation If the microscope is not properly aligned good results will not be obtained I Start Up This procedure is when the micro ...